Identification¶
Model: ECG1200G
12-channel Electrocardiograph is such a kind of electrocardiograph, which gathers 12-lead ECG signals simultaneously and prints out the ECG waveforms with thermal printing system.
Documentation¶
- Service Manual: https://tickets.chezo.africa/attachments/11
Issues¶
The device shows no signal when the electrode are on.¶
Description
- The device is switched on
- The cables are connected at the device
- At least, the big clips cables are connected to a living person (hand+foot)
- The device is set to measure something
- No changing signal is shown.
Investigation Results
- The Sensor board is powered when the measurement is enabled.
- Electricity can be measured at the connector to the cables.
- All cables are checked using a Multimeter and showing a resistance less than 10 Ohm (as specified)
Images¶
Files
C( Updated by Christoph (Obruni) 4 months ago
C( Updated by Christoph (Obruni) 4 months ago
- File PXL_20250815_174711030.MP.jpg PXL_20250815_174711030.MP.jpg added
- Description updated (diff)
C( Updated by Christoph (Obruni) 4 months ago
- Status changed from Running to Broken
RA Updated by Redmine Admin 4 months ago
- Subject changed from ECG1200G Electrocardiograph to Electrocardiograph Contec ECG1200G
- Description updated (diff)